top of page

Most Graph-02

Respiratory Diagnosis without Patient's Cooperation

Most Graph-02

The Chest MostGraph-02 is a sophisticated medical device designed for the total measurement of respiratory resistance and reactance using the Forced Oscillatory Technique (FOT). Here are the key details:

  • Measurement Technique: Utilizes the Forced Oscillatory Technique (FOT) for assessing total respiratory resistance and reactance.

  • Testing: Quick and simple testing by quiet breathing, typically completed in 20 seconds.

  • Results Interpretation: Provides graphic 3D interpretation of the results, making it easy for patients to understand.

  • Device Features:
    Windows-based standalone device with a touchscreen monitor.
    Spirometer option available.

  • Specifications:
    Flow Detection: Lilly type pneumotach sensor.
    Flow Range: 0 to ±2.6 L/s.
    Pressure Range: 0 to ±5.00 cmH2O.
    Pressure Accuracy: ±3.0% (at 2 cmH2O/L/s).
    Pulse Wave Frequency Range: 4 to 36 Hz, at every 4 Hz.
    Time Resolution: 0.25 seconds.
    Impedance Interpolation: Spline interval of 0.25 seconds.
    Noise Wave Frequency Range: 4 to 36 Hz, at every 2 Hz.
    Power Supply: AC 100 – 240V, 50/60Hz.
    Dimensions: W35 × D32 × H27 cm (Main Unit).
    Weight: Approximately 11 kg.

  • Measurement Items:
    FOT: R5, R20, R5-R20, X5, Fres, ALX.
    Spirometry: SVC, FVC, TV, FEV0.5, FEV1, FEV3, FEV6, FEV0.5/FVC, FEV1/FVC, FEV1/SVC, FEV3/FVC, FEV6/FVC, FEV1/SVCpr, MMEF, PEF, FEF25, FEF50, FEF75, FEF90, MMEF/FVC, FEF50/FEF75, FEF75-85, FEF200-1200, PEF time, FET, ExtrapV, ExtrapV%, FIVC, FIV0.5, FIV1, FIV1/FVC, FIV1/FIVC, PIF, FIF50, FEF50/FIF50, FIF50/FEF50, MVV43, FVC+FEV1, MVV2.

bottom of page